McCrometer - US
Join Our Email List 1-800-220-2279 

FPI Mag Plus 3000

FPI Mag Plus 3000
Series: FPI-MAG-PLUS-3000
Unit Price Contact McCrometer

Specifications

Accessories: - Sensor insertion tool
- Extension to hardware clearance
- Additional sensor cable up to 475’ (500’ max for model 395 and 200’ max for model 394)
Accuracy: AC or DC power: ±1% of measured value ±0.006 ft/s (± 0.0018 m/s)
Battery power: ±1% of measured value ±0.006 ft/s (± 0.0018 m/s)
Reverse Flow: ±1% of measured value ±0.006 ft/s (± 0.0018 m/s)
Body material: Fusion bonded epoxy (NSF 61 approved) coated 316 stainless steel
Certifications: Standard model:
- ISO 9001:2015 certified quality management system
- NSF/ANSI/CAN 61 & NSF/ANSI 372
- Certified by MET to UL 61010-1

HL model:
- ISO 9001:2015 certified quality management system
- NSF/ANSI/CAN 61 & NSF/ANSI 372
- Certified by MET to UL 61010-1 and MET C22.2 No. 61010-1-04
- - Class I, Division 2, Groups A-D, T5
- - Class I, Zone 2, IIC T5
Converter: ProComm GO Transmitter
Directional Response: Directional / Bidirectional

Direction Measurement:
- 395 sensor: Forward flow measurement and reverse flow indication
- 394 sensor: Bidirectional flow measurement
Electrical Connections: - Quick Connect
- Compression gland seals
Electrode: 316 Stainless Steel
IP Rating: Standard model: Quick Connect (IP68), Compression gland seals (IP68)
HL model: Quick Connect (IP67), Compression gland seals (IP67)
Linearity: 0.3 % of Range
Line size: Up to 130”
Material: - Stainless steel body
- 316 stainless steel electrodes
- NSF approved fusion bonded epoxy coating
- 2” stainless steel ball valve (minimum of 1-7/8” port I.D.)
- 2” x close stainless steel nipple
- Coating: Fusion bonded epoxy (NSF 61 approved) coated 316 stainless steel
- Compression Seal: Silicone Rubber
Measurement Method: Volumetric flow in filled flow conduits 4” (100 mm) to 138” (3,500 mm) utilizing insertable electromagnetic averaging sensor. Flow indication in English Standard or Metric units.
Mounting: Converter: Remote mount
Operating Conditions: Standard model: 10 to 60 °C (14 to 140 °F) up to 250 PSI
HT (High Temperature) option: 10 to 93°C (14 to 200°F) up to 250 PSI
Options: - Hastelloy® electrodes
- Annual verification / calibration
- Stainless steel ID tag
Power: ProComm Go transmitter can use AC power, DC power, and battery
Repeatability: 0.2% of reading 
Requirements: Pipe run requirements for the FPI Mag sensor can vary, depending on the disturbances in the pipe run. See FPI Mag 3000 flow meter manual 30126-03, step 3 of the installation procedure for a table of upstream and downstream straight-pipe run recommendations.
Seal material: 316 Stainless Steel
Sensor Cable: Standard: 25’/7.6 m McCrometer supplied submersible cable with each remote mount unit.
Optional: Up to 500’/152.4 m, or 25’/7.6 m max for battery powered.
Quick connect: Available in standard cable lengths:
- Feet: 25, 50, 75, 100, 125, 150, 175, 200, 500
- Meters: 7.6, 15.25. 22.5, 30.5, 38.1, 45.75, 53.3, 61, 152.4

Custom cable lengths at additional cost.
Sensor Immersion Depth: - With standard quick connect: 1.8 m (6 ft.)
- With optional strain relief cable: 9 m (30 ft.)
Storage Conditions: -15 to 60°C (5 to 140 °F), 15 to 93°C (5 to 200° F) for High Temperature Option
Note regarding storage: During freezing conditions and when meter is not in use, sensor must be removed from pipe and stored in dry conditions.
Note: Damage to the sensor caused by allowing the sensor freeze in the pipe is not covered by the warranty.
Velocity Measurement Range: See FPI Mag Specification Sheet, 30126-15, for flow meter pipe sizes and flow ranges with ProComm GO Transmitter.
Warranty: 24 months